CR 85-X

Key benefits

  • 10 Cassette Input and output buffer for maximized productivity
  • No waiting times for maximized productivity
  • For a broad range of clinical applications
  • Optimal accessibility

CR 85-X is a multi-plate, multi-user digitizer featuring a unique drop-and-go buffer that eliminates waiting times and maximizes productivity.

CR 85-X is also a multi-application digitizer, benefiting from three different image resolution modes and maximizing productivity for the complete clinical application range.

Suitable for following market segments

  • General radiography
  • Orthopaedics and Extremities
  • Mammography (Note: mammography use is not available in the U.S.)

Availability

Worldwide

The CR 85-X Digitizer is designed as a high throughput centralized system. It features a unique drop-and-go buffer that eliminates waiting times and maximizes productivity.

Highest productivity

The cassette buffer eliminates waiting times and allows for a continuous workflow within the department. Zero-button operation with automated cassette handling makes CR 85-X a highly productive and user-friendly system with a throughput of up to 112 plates an hour, depending on the cassette size and application.


No waiting

The CR 85-X Digitizer requires no manual interaction and all the user has to do is to deposit the cassettes. The digitizer automatically takes up to 10 cassettes from the input buffer, reads the demographic data from the memory on the cassette, scans the imaging plate, digitizes the image and returns the cassette to the output buffer for new exposures.


Full data

CR 85-X reads imaging plates at a standard resolution of 6 pixels/mm. The high resolution mode of 10 pixels/mm is available for all image plate sizes. The maximum resolution mode of 20 pixels/mm is available for dedicated 18 x 24 cm and 24 x 30 cm extremities cassettes and plates.


Compact footprint and optimal accessibility

CR 85-X occupies a very small floor space and at the same time provides unhindered access to multiple users, one at the input and one at the output buffer, resulting in a smooth flow of operations. This concept makes CR 85-X an excellent solution for centralized CR.


Broad range of applications

In combination with the application-specific plates and cassettes, CR 85-X supports a broad range of applications:
General Radiography
Orthopaedics - Extremities
Mammography (Note: Mammography use is not available in the U.S.)

GENERAL

Cassette buffer capacity and performance
10 cassettes of mixed sizes, both in input and output buffer
Throughput: up to 112 plates/hr. (depending on size and application)

LCD display
Machine status and error conditions

Greyscale resolution
Data acquisition: 12 bits/pixel
Output to processor: 12 bits/pixel

Dimensions and weight
W x D x H: 84 x 115 x 142 cm (33 x 45 x 56")
At foot: 84 cm (33")
At buffer: 142 cm (56")
Weight: Approx. 320 kg (705.48 lbs)

Power
50/60 Hz single phase
240V +10%, max. fuse 16A
230V ±10%, max. fuse 16A
208V ±10%, max. fuse 15A (e.g., USA)
200V ±10%, max. fuse 15A (e.g., Japan)

Environmental conditions
Temperature: 20˚- 30˚C (68° - 86° F)
Humidity: 15 - 75% RH
Magnetic fields: max. 12.60μT
Rate of change of temperature: 0.5° C/minute (0.9° F/minute)

Environmental effects
Noise level: max. 65 dB (A)
Heat dissipation: standby 350 W, max. 2000 W

SAFETY

Approvals
TüV, UL, cUL, CE

Transport details
Temperature: -25 to +55° C (-4 to 131° F), -25° C for max. 72 hours, +55° C for max. 96 hours
Humidity: 5 - 95% RH

CR 85-X
MUSICA2

Advanced Image Processing Technology

MUSICA² More