DX-M

Digitizer that supports both standard phosphor plates and needle-based detectors, for the next generation in computed radiography (CR). DX-M is ideal for a broad range of applications including mammography and general radiography.

Key benefits

  • Mammography image quality, with potential dose reduction.
  • Smoother, more productive workflow, with multi-cassette drop-and-go buffer.
  • For both centralized and decentralized environments.
  • Small, convenient footprint.

The DX-M digitizer unites the superb image quality necessary for mammography, with high throughput, delivered by a drop-and-go buffer-based workflow.

The combination of needle-based detectors and standard phosphor plates with specific cassettes and image resolution mode makes the DX-M ideal for a broad range of applications, including general radiography, mammography, orthopedics/extremities, dental, pediatrics/neonatal and full leg/full spine.

DX-M is a CR solution with needle-based detectors to reach the high level quality demanded by mammography. One that has the potential for dose reduction for all studies - particularly neonatology and pediatrics, where dose reduction is crucial. That combines superior image quality with high throughput, delivered by a unique five cassette drop-and-go buffer, and a very fast preview - all housed in a compact system.

Suitable for following market segments

General radiography
Digital Mammography
Orthopedics - extremities
Dental
Pediatrics and neonatal
Full Leg / Full Spine

Availability

Worldwide, except for US.

Three image resolution modes

It offers three different image resolution modes: 50 μm pixel pitch (20 pixels/mm), 100 μm pixel pitch (10 pixels/mm) and 150 μm pixel pitch (6.7 pixels/mm).


Maximum productivity and smooth workflow

The drop-and-go buffer and fast preview eliminate waiting times and facilitate a continuous workflow. The five-cassette drop-and-go buffer can handle a mix of different sizes of both needle-based detectors and standard phosphor plates. The automatic cassette handling makes DX-M highly productive and user-friendly. As a central digitizer in the radiography department, the DX-M can serve multiple rooms. With its small footprint, it can fit into the tightest spaces, including the X-ray room or even a narrow corridor.


The right choice

To eliminate any confusion, needle-based detector cassettes are gray, while standard phosphor plate cassettes are orange, so that there is no chance of the user making a mistake when selecting the desired cassette.


Embedded patient ID

Each plate has an embedded memory that stores the data entered during identification by no-touch radiofrequency tagging. Identification data and images are linked from the beginning throughout the entire digital processing system.

Drop-and-go cassette buffer

5 cassettes of mixed sizes input buffer
5 cassettes of mixed sizes output buffer.

Throughput


35 x 43 cm (14 x 17") = approx. 83 plates/hour.

Display for status and error indication


LCD touchscreen
LED status indicator

Grayscale resolution

Output to processor: 16 bits/pixel square root compressed

Dimensions and weight


Covered floor space (W x D x H): 66 x 51 x 123 cm (26 x 20 x 48.4")
Output buffer included (W x D x H): 115 x 51 x 123 cm (45.3 x 20 x 48.4")
Weight: approx.: 180 kg (397 lbs)

Configuration requirements


NX
ID tablet
CR HD5.0 detectors and cassettes
CR MD4.0R plates and cassettes
CR HM5.0 detectors and cassettes
CR MM3.0R plates and cassettes

Power


220-240 V/50-60 Hz standby 87 W, peak 590 W, fuse 16A
120 V/60 Hz (USA) standby 92 W, peak 621 W, fuse 15A
100 V/60 Hz (Japan) standby 92 W, peak 621 W, fuse 15A

Environmental conditions

Temperature: 15-30°C (59-86°F)
Temperature when used with CR HM5.0 Detector: 20-30°C
Humidity: 15-75% RH
EMC compliant with IEC 60601-1-2
Rate of change of temperature: 0.5°C/minute (0.9°F)

Environmental effects


Noise level: max. 65 dB (A)
Heat dissipation: standby 92 W, continuous operation 242 W

SAFETY

Approvals


ETL classified CUS, CE

Transport details


Temperature: -25 to +55°C (-4 to 131°F), -25°C for max. 72 hours, +55°C for max. 96 hours
Humidity: 5-95% RH

DX-M
MUSICA2

Advanced Image Processing Technology

MUSICA²